This shows you the differences between two versions of the page.
| Next revision | Previous revision | ||
|
private:writeups:plasmaimpedanceprobes [2020/07/01 17:30] admin created |
private:writeups:plasmaimpedanceprobes [2020/07/02 14:17] (current) admin |
||
|---|---|---|---|
| Line 3: | Line 3: | ||
| === What it does === | === What it does === | ||
| - | === Advantages/disadvantages | + | Measures |
| + | - Plasma frequency [Blackwell2005] | ||
| + | - Electron density (by measuring plasma frequency) [Blackwell2005] | ||
| + | - Plasma potential | ||
| + | - Collision frequency | ||
| + | |||
| + | === Methods === | ||
| + | |||
| + | PIPs can be operated in a number of methods | ||
| + | |||
| + | - Actively driven RF sweeps [Blackwell2005]. | ||
| + | - Plasma oscillation (excite a two stream instability) [Blackwell2005]. | ||
| + | |||
| + | |||
| + | === Advantages | ||
| + | |||
| + | * [Blackwell2005] An improvement over Langmuir probes because contamination of the probe has a larger impact on its measurements than on the PIP. | ||
| + | * Better resolution and accuracy than traditional Langmuir probes [citation] | ||
| + | |||
| + | === Disadvantages === | ||
| + | |||
| + | * More complicated. | ||
| + | * (Check this) Cannot be used in real-time because the network analyzer data can only be post-processed. | ||
| + | |||
| + | === Sources of error === | ||
| + | |||
| + | * [Blackwell2005] recommends " | ||
| + | * [Blackwell2005] Nearby objects to the probe reflected radiation | ||
| + | * [Blackwell2005] At frequencies related to certain wavelengths, | ||
| === Types of electrodes used === | === Types of electrodes used === | ||
| Line 11: | Line 39: | ||
| * Monopole antenna [Bilen1999] | * Monopole antenna [Bilen1999] | ||
| + | |||
| + | === [Blackwell2005] === | ||
| + | |||
| + | Blackwell2005 discusses a PIP in a cold (cold ions) and unmagnetized plasma using a spherical probes. | ||
| + | |||
| + | Derivation of Eq. 2. Also citing Piel2009. | ||
| + | |||
| + | {{ : | ||
| + | |||
| + | === Calibration === | ||
| + | |||
| + | A calibration is common to remove the impedance of the non-plasma components (coax feedlines, network analyzer, etc). | ||
| + | * [Blackwell2005] does this by attaching a 50ohm load and then a short at the end of the feedline with the other end attached to a network analyzer. | ||