 === What it does === === What it does ===
  
-=== Advantages/disadvantages ===+Measures 
 +  - Plasma frequency [Blackwell2005] 
 +  - Electron density (by measuring plasma frequency) [Blackwell2005] 
 +  - Plasma potential 
 +  - Collision frequency 
 + 
 +=== Methods === 
 + 
 +PIPs can be operated in a number of methods 
 + 
 +  - Actively driven RF sweeps [Blackwell2005].  Relate the peaks and zero intercepts of Re(Z) and Im(Z) to circuit models and dispersion relationships to get the plasma frequency. 
 +  - Plasma oscillation (excite a two stream instability) [Blackwell2005].  Drive weak beam of electrons into the plasma to excite a two stream instability.  Under correct conditions, its frequency is close to the plasma frequency.  An FFT of passive PIP measurements reveals the frequency.   
 + 
 + 
 +=== Advantages === 
 + 
 +  * [Blackwell2005] An improvement over Langmuir probes because contamination of the probe has a larger impact on its measurements than on the PIP. 
 +  * Better resolution and accuracy than traditional Langmuir probes [citation] 
 + 
 +=== Disadvantages === 
 + 
 +  * More complicated.  Requires careful setup/calibration and the use of an expensive network analyzer. 
 +  * (Check this) Cannot be used in real-time because the network analyzer data can only be post-processed.   
 + 
 +=== Sources of error === 
 + 
 +  * [Blackwell2005] recommends "choosing sufficiently shielded coax cables so that the impedance characteristics are not changed as cables are moved around during the experiment. " 
 +  * [Blackwell2005] Nearby objects to the probe reflected radiation 
 +  * [Blackwell2005] At frequencies related to certain wavelengths, standing waves can form in the chamber.
  
 === Types of electrodes used === === Types of electrodes used ===
Line 11: Line 39:
   * Monopole antenna [Bilen1999]   * Monopole antenna [Bilen1999]
  
 +
 +=== [Blackwell2005] ===
 +
 +Blackwell2005 discusses a PIP in a cold (cold ions) and unmagnetized plasma using a spherical probes.  Provides a derivation of the circuit model (I reproduce part of it below) that captures the impedance of the sheath and the plasma.   
 +
 +Derivation of Eq. 2.   Also citing Piel2009.  
 +
 +{{ :private:writeups:img_20200701_140114.png?nolink&600 |}}
 +
 +=== Calibration ===
 +
 +A calibration is common to remove the impedance of the non-plasma components (coax feedlines, network analyzer, etc).  
 +  * [Blackwell2005] does this by attaching a 50ohm load and then a short at the end of the feedline with the other end attached to a network analyzer.  "the calibration consists of programming the analyzer to add phase shifts to the value of G to cancel out the phase shift from the cable."